Input |
• Rated input voltage |
85÷264Vac |
|
• Line frequency |
50/60Hz |
|
• Power factor |
0.9 typ. |
|
• Leakage current to GND (max) |
<1.5mA |
|
• Input protection |
• Fuse 1x T 5A • Mains filter • Start-up peak current limitation |
|
• Output voltage temp drift |
0.01%/°C |
Output |
• Output power (max) |
290W |
|
• Hold up time |
20msec typ. |
|
• Output protections |
• Short circuit on all outputs • Overload on all outputs • Overvoltage on outputs: A,B,C,D,E • Undervoltage on outputs: A,B • Thermal shut-down (85°C typ.) |
|
• Output signals |
• INIT low active TTL compatible • THERM. FAIL low active TTL compatible • POWER FAIL low active TTL compatible |
General Features |
• DC-DC / AC-DC |
AC - DC |
|
• Efficiency |
75% typ. |
|
• Dieletric withstand voltage |
• Input - Output :1750Vac • Input -P.E.: 1750Vac |
|
• Cooling |
Forced ventilation |
|
• Operating temperature |
-15°C to 40°C |
|
• Switching frequency |
60KHz typ. |
|
• Comply with |
• EN 50081-1 • EN 50082-1 • EN 61010 • CSA 1010 • UL 1262 • UL 3101 |
|
• Application fiels |
Medical application |